Tidal Volume
Delivered tidal volumes with the healthy lung model as a function of set tidal volume on the ASV in volume control mode connected to a passive test lung. For comparison, a dashed line depicts perfect linearity. The EURS guidance recommends tidal volume delivery range from 350 to 450 ml

Respiratory Rate
Measured respiratory rate in breaths per minute versus set respiratory rate for the ASV connected to a passive lung. For comparison, a dashed line depicts perfect linearity. The EURS required respiratory rate range is 10 to 30 bpm with a 2-bpm tolerance